Pixel Response Effects On CCD Camera Gain Calibration
|
|
|
Introduction
|
The gain of a CCD camera is the conversion between the number of electrons ("e-")
recorded by the CCD and the number of digital units ("counts") contained in the
CCD image. It is useful to know this conversion for evaluating the performance
of the CCD camera. Since quantities in the CCD image can only be measured in
units of counts, knowing the gain permits the calculation of quantities such as
readout noise and full well capacity in the fundamental units of electrons.
The gain value is required by some types of image Deconvolution such as Maximum
Entropy since, in order to do properly the statistical part of the calculation,
the processing needs to convert the image into units of electrons.
Calibrating the gain is also useful for detecting electronic problems in a CCD camera,
including gain change at high or low signal level, and the existence of unexpected
noise sources.
This White Paper develops the mathematical theory behind the gain calculation and
shows how the mathematics suggests ways to measure the gain accurately. This note
does not address the issues of basic image processing or CCD camera operation,
and a basic understanding of CCD bias, dark and flat field correction is assumed.
|
|
|
|
Overview
|
The gain value is set by the electronics that read out the CCD chip. Gain is
expressed in units of electrons per count. For example, a gain of 1.8 e-/count
means that the camera produces 1 count for every 1.8 recorded electrons. Of course,
we cannot split electrons into fractional parts, as in the case for a gain of
1.8 e-/count. What this number means is that 4/5 of the time 1 count is produced
from 2 electrons, and 1/5 of the time 1 count is produced from 1 electron. This
number is an average conversion ratio, based on changing large numbers of electrons
into large numbers of counts. Note: This use of the term "gain" is in the opposite
sense to the way a circuit designer would use the term since, in electronic design,
gain is considered to be an increase in the number of output units compared
with the number of input units.
It is important to note that every measurement you make in a CCD image uses
units of counts. Since one camera may use a different gain than another camera,
count units do not provide a straightforward comparison to be made. For example,
suppose two cameras each record 24 electrons in a certain pixel. If the gain of
the first camera is 2.0 and the gain of the second camera is 8.0, the same pixel
would measure 12 counts in the image from the first camera and 3 counts in the
image from the second camera. Without knowing the gain, comparing 12 counts
against 3 counts is pretty meaningless.
Before a camera is assembled, the manufacturer can use the nominal tolerances of
the electronic components to estimate the gain to within some level of uncertainty.
This calculation is based on resistor values used in the gain stage of the CCD readout
electronics. However, since the actual resistance is subject to component tolerances,
the gain of the assembled camera may be quite different from this estimate. The actual
gain can only be determined by actual performance in a gain calibration test. In addition,
manufacturers sometimes do not perform an adequate gain measurement. Because of
these issues, it is not unusual to find that the gain of a CCD camera differs
substantially from the value quoted by the manufacturer.
|
|
|
|
Background
|
The signal recorded by a CCD and its conversion from units of electrons to counts
can be mathematically described in a straightforward way. Understanding the mathematics
validates the gain calculation technique described in the next section, and it shows
why simpler techniques fail to give the correct answer.
This derivation uses the concepts of "signal" and "noise". CCD performance is usually
described in terms of signal to noise ratio, or "S/N", but we shall deal with them
separately here. The signal is defined as the quantity of information you measure
in the image—in other words, the signal is the number of electrons recorded by the CCD
or the number of counts present in the CCD image. The noise is the uncertainty in the signal.
Since the photons recorded by the CCD arrive in random packets (courtesy of nature),
observing the same source many times records a different number of electrons every time. This
variation is a random error, or "noise" that is added to the true signal. You
measure the gain of the CCD by comparing the signal level to the amount of
variation in the signal. This works because the relationship between counts and
electrons is different for the signal and the variance. There are two
ways to make this measurement:
-
Measure the signal and variation within the same region of pixels at many intensity levels.
-
Measure the signal and variation in a single pixel at many intensity levels.
Both of these methods are detailed in section 6. They have the
same mathematical foundation.
To derive the relationship between signal and variance in a CCD
image, let us define the following quantities:
|
|
|
| SC |
The signal measured in count units in the CCD image. |
| SE |
The signal recorded in electron units by the CCD chip. This quantity is unknown. |
| NC |
The total noise measured in count units in the CCD image. |
| NE |
The total noise in terms of recorded electrons. This quantity is unknown. |
| g |
The gain, in units of electrons per count. This will be calculated. |
| RE |
The readout noise of the CCD, in [electrons]. This quantity is unknown. |
| sE |
The photon noise in the signal NE. |
| sO |
An additional noise source in the image. This is described below. |
|
We need an equation to relate the number of electrons, which is unknown, to
quantities we measure in the CCD image in units of counts. The signals and
noises are simply related through the gain factor as:
and

These can be inverted to give:
and

The noise is contributed by various sources. We consider these to be readout
noise, RE, photon noise attributable to the nature of light,
σE and some additional noise, σo,C,
which will be shown to be important in in the following Section. Remembering
that the different noise sources are independent of each other, they add in quadrature.
This means that they add as the square their noise values. If we could measure
the total noise in units of electrons, the various noise sources would combine
in the following way:

The random arrival rate of photons controls the photon noise, σE.
Photon noise obeys the laws of Poissonian statistics, which makes the square of the noise equal to
the signal, or
.
Therefore, we can make the following substitution:
.
Knowing how the gain relates units of electrons and counts, we can modify this
equation to read as follows:

which then gives

We can rearrange this to get the final equation:

This is the equation of a line in which
is the y axis, SC is the x axis, and the slope is 1/g. The extra terms
are grouped together for the time being. Below, they will be separated, as the extra
noise term has a profound effect on the method we use to measure gain. A better way
to apply this equation is to plot our measurements with SC
as the y axis and
as the x axis, as this gives the gain directly as the slope. Theoretically, at least,
one could also calculate the readout noise,
,
from the point where the line hits the y axis at SC= 0.
Knowing the gain then allows this to be converted to a Readout Noise in the
standard units of electrons. However, finding the intercept of the line is
not a good method, because the readout noise is a relatively small quantity
and the exact path where the line passes through the y axis is subject to
much uncertainty.
With the mathematics in place, we are now ready to calculate the gain. So far,
I have ignored the "extra noise term", σo,C.
In the next 2 sections, I will describe the nature of the extra noise term and show
how it affects the way we measure the gain of a CCD camera.
|
|
|
|
Crude Estimation of the Gain
|
In the previous section we derived the complete equation that
relates the signal and the noise you measure in a CCD image. One popular method
for measuring the gain is very simple, but it is not based on the full equation
I have derived above. The simple can be described as follows:
-
Obtain images at different signal levels and subtract the bias
from them. This is necessary because the bias level adds to the measured signal
but does not contribute noise.
-
Measure the signal and noise in each image. The mean and standard deviation of a
region of pixels give these quantities. Square the noise value to get a variance
at each signal level.
-
For each image, plot Signal on the y axis against Variance on the x axis.
-
Find the slope of a line through the points. The gain equals the slope.
Is measuring the gain actually this simple? Well, yes and no.
If we actually make the measurement over a substantial range of signal, the data
points will follow a curve rather than a line. Using the present method we will
always measure a slope that is too shallow, and with it we will always
underestimate the gain. Using only low signal levels, this method can give a
gain value that is at least "in the ballpark" of the true value. At low signal
levels, the curvature is not apparent, though present. However, the data points
have some amount of scatter themselves, and without a long baseline of signal,
the slope might not be well determined. The curvature in the Signal - Variance
plot is caused by the extra noise term which this simple method neglects.
The following factors affect the amount of curvature we obtain:
- The color of the light source. Blue light is worse
because CCD’s show the greatest surface irregularity at shorter
wavelengths. These irregularities are described in Section 5.
-
The fabrication technology of the CCD chip. These issues
determine the relative strength of the effects described in item 1.
-
The uniformity of illumination on the CCD chip. If the Illumination is not
uniform, then the sloping count level inside the pixel region used to measure it
inflates the measured standard deviation.
Fortunately, we can obtain the proper value by doing just a bit
more work. We need to change the experiment in a way that makes the data plot as
a straight line. We have to devise a way to account for the extra noise term,
σo,C. If σo,C were a constant value we could
combine it with the constant readout noise. We have not talked in detail about
readout noise, but we have assumed that it merges together all constant noise
sources that do not change with the signal level.
|
|
|
|
The Extra Noise Term in the Signal-Variance Relationship
|
The mysterious extra noise term, σo,C, is attributable
to pixel-to-pixel variations in the sensitivity of the CCD, known as the
flat field effect. The flat field effect produces a pattern of
apparently "random" scatter in a CCD image. Even an exposure with infinite
signal to noise ratio ("S/N") shows the flat field pattern. Despite its
appearance, the pattern is not actually random because it repeats from one
image to another. Changing the colour of the light source changes the
details of the pattern, but the pattern remains the same for all images
exposed to light of the same spectral makeup. The importance of this effect
is that, although the flat field variation is not a true noise, unless it
is removed from the image it contributes to the noise you actually measure.
We need to characterize the noise contributed by the flat field
pattern in order to determine its effect on the variance we measure in the
image. This turns out to be quite simple: Since the flat field pattern is a
fixed percentage of the signal, the standard deviation, or "noise" you measure
from it is always proportional to the signal. For example, a pixel might be 1%
less sensitive than its left neighbor, but 3% less sensitive than its right
neighbor. Therefore, exposing this pixel at the 100 count level produces the
following 3 signals: 101, 100, 103. However, exposing at the 10,000 count level
gives these results: 10,100, 10,000, 10,300. The standard deviation for these 3
pixels is σo,C = 2.333 counts for the low signal case but is
σo,C = 233.3 counts for the high signal case. Thus the standard
deviation is 100 times larger when the signal is also 100 times larger. We can
express this proportionality between the flat field "noise" and the signal
level in a simple mathematical way:
σo,C = kSC
In the present example, we have k=0.02333. Substituting this expression
for the flat field variation into our master equation, we get the following result:

With a simple rearrangement of the terms, this reveals a nice quadratic function of signal:

When plotted with the Signal on the x axis, this equation describes a
parabola that opens upward. Since the Signal - Variance plot is actually
plotted with Signal on the y axis, we need to invert this equation to solve for
SC:

This final equation describes the classic Signal - Variance plot.
In this form, the equation describes a family of horizontal parabolas that
open toward the right. The strength of the flat field variation,
k, determines the curvature. When k = 0, the curvature goes away
and it gives the straight line relationship we desire. The curvature to the
right of the line means that the stronger the flat field pattern, the more the
variance is inflated at a given signal level. This result shows that it is
impossible to accurately determine the gain from a Signal - Variance plot
unless we know one of two things: Either 1) we know the value of k, or 2)
we setup our measurements to avoid flat field effects. Option 2 is the correct
strategy. Essentially, the weakness of the method described in Section 4 is that
it assumes that a straight line relationship exists but ignores flat field
effects.
To illustrate the effect of flat field variations, mathematical
models were constructed using the equation above with parameters typical of
commonly available CCD cameras. These include readout noise RE
= 15e- and gain g = 2.0 e- / Count. Three models were constructed with
flat field parameters k = 0, k = 0.005, and k = 0.01. Flat
field variations of this order are not uncommon. These models are shown in the
figure below.
|
|
|
 |
|
Increasing values of k correspond to progressively
larger flat field irregularities in the CCD chip. The amplitude of flat field
effects, k, tends to increase with shorter wavelength, particularly with
thinned CCD's (this is why Section 4 recommends using a redder light source to
illuminate the CCD). The flat field pattern is present in every image
exposed to light. Clearly, it can be seen from the models that if one simply
obtains images at different signal levels and measures the variance in them,
then fitting a line through any part of the curve yields a slope lower
than its true value. Thus the simple method of section 4 always underestimates
the gain.
The best strategy for doing the Signal - Variance method is to
find a way to produce a straight line by properly compensating for flat field
effects. This is important by the "virtue of straightness": Deviation from a
straight line is completely unambiguous and easy to detect. It avoids the
issue of how much curvature is attributable to what cause. The electronic design
of a CCD camera is quite complex, and problems can occur, such as gain change at
different signal levels or unexplained extra noise at high or low signal levels.
Using a "robust" method for calculating gain, any significant deviation from a
line is a diagnostic of possible problems in the camera electronics. Two such
methods are described in the following section.
|
|
|
|
Robust Methods for Measuring Gain
|
In previous sections, the so-called simple method of estimating
the gain was shown to be an oversimplification. Specifically, it produces a
Signal - Variance plot with a curved relationship resulting from flat field
effects. This section presents two robust methods that correct the flat field
effects in the Signal - Variance relationship to yield the desired straight-line
relationship. This permits an accurate gain value to be calculated. Adjusting
the method to remove flat field effects is a better strategy than either to
attempt to use a low signal level where flat field effects are believed not to
be important or to attempt to measure and compensate for the flat field
parameter k.
When applying the robust methods described below, one must consider some procedural
issues that apply to both:
Both methods measure sets of 2 or more images at each signal
level. An image set is defined as 2 or more successive images taken under the
same illumination conditions. To obtain various signal levels, it is better to
change the intensity received by the CCD than to change the exposure time. This
may be achieved either by varying the light source intensity or by altering the
amount of light passing into the camera. The illumination received by the CCD
should not vary too much within a set of images, but it does not have to be
truly constant.
-
Cool the CCD camera to reduce the dark current to as low as
possible. This prevents you from having to subtract dark frames from the images
(doing so adds noise, which adversely affects the noise measurements at low
signal level). In addition, if the bias varies from one frame to another, be
sure to subtract a bias value from every image.
-
The CCD should be illuminated the same way for all images
within a set. Irregularities in illumination within a set are
automatically removed by the image processing methods used in the calibration.
It does not matter if the illumination pattern changes when you change the
intensity level for a different image set.
-
Within an image set, variation in the light intensity is
corrected by normalizing the images so that they have the same average signal
within the same pixel region. The process of normalizing multiplies the image by
an appropriate constant value so that its mean value within the pixel region
matches that of other images in the same set. Multiplying by a constant value
does not affect the signal to noise ratio or the flat field structure of the
image.
-
Do not estimate the CCD camera's readout noise by calculating
the noise value at zero signal. This is the square root of the variance where
the gain line intercepts the y axis. Especially do not use this value if bias is
not subtracted from every frame. To calculate the readout noise, use the "Two
Bias" method and apply the gain value determined from this test. In the Two Bias
Method, 2 bias frames are taken in succession and then subtracted from each
other. Measure the standard deviation inside a region of, say 100x100 pixels and
divide by 1.4142. This gives the readout noise in units of counts. Multiply this
by the gain factor to get the Readout Noise in units of electrons. If bias
frames are not available, cool the camera and obtain two dark frames of minimum
exposure, then apply the Two Bias Method to them.
|
|
|
Method 1: Correct the flat field effects at each signal level
In this strategy, the flat field effects are removed by subtracting one
image from another at each signal level. Here is the recipe:
For each intensity level, do the following:
-
Obtain 2 images in succession at the same light level. Call
these images A and B.
-
Subtract the bias level from both images. Keep the exposure
short so that the dark current is negligibly small. If the dark
current is large, you should also remove it from both frames.
-
Measure the mean signal level S in a region of pixels on
images A and B. Call these mean signals SA and
SB. It is best if the bounds of the region change as little as
possible from one image to the next. The region might be as small as 50x50 to
100x100 pixels but should not contain obvious defects such as cosmic ray hits,
dead pixels, etc.
-
Calculate the ratio of the mean signal levels as r =
SA / SB.
-
Multiply image B by the number r. This corrects image B to the same signal
level as image A without affecting its noise structure or flat field variation.
-
Subtract image B from image A. The flat field effects present
in both images should be cancelled to within the random errors.
-
Measure the standard deviation over the same pixel region you used in step 3.
Square this number to get the Variance. In addition, divide the resulting variance
by 2.0 to correct for the fact that the variance is doubled when you subtract one
similar image from another.
-
Use the Signal from step 3 and the Variance from step 7 to add
a data point to your Signal - Variance plot.
-
Change the light intensity and repeat steps 1 through 8.
|
|
|
Method 2: Avoid flat field effects using one pixel in many images.
This strategy avoids the flat field variation by considering how a single pixel
varies among many images. Since the variance is calculated from a single pixel
many times, rather than from a collection of different pixels, there is no flat
field variation.
To calculate the variance at a given signal level, you obtain many frames, measure
the same pixel in each frame, and calculate the variance among this set of
values. One problem with this method is that the variance itself is subject to
random errors and is only an estimate of the true value. To obtain a reliable
variance, you must use 100’s of images at each intensity level. This is
completely analogous to measuring the variance over a moderate sized pixel
region in Method A; in both methods, using many pixels to compute the variance
gives a more statistically sound value. Another limitation of this method is
that it either requires a perfectly stable light source or you have to
compensate for light source variation by adjusting each image to the same
average signal level before measuring its pixel. Altogether, the method requires
a large number of images and a lot of processing. For this reason, Method A is
preferred. In any case, here is the recipe:
Select a pixel to measure at the same location in every image. Always measure
the same pixel in every image at every signal level.
For each intensity level, do the following:
-
Obtain at least 100 images in succession at the same light level. Call
the first image A and the remaining images i. Since you are
interested in a single pixel, the images may be small, of order 100x100 pixels.
-
Subtract the bias level from each image. Keep the exposure short so that the
dark current is negligibly small. If the dark current is large, you should
also remove it from every frame.
-
Measure the mean signal level S in a rectangular region
of pixels on image A. Measure the same quantity in each of the remaining images.
The measuring region might be as small as 50x50 to 100x100 pixels and should be
centered on the brightest part of the image.
-
For each image Si other than the first,
calculate the ratio of its mean signal level to that of image A. This gives a
number for each image, ri = SA /
Si.
-
Multiply each image i by the number ri. This corrects
each image to the same average intensity as image A.
-
Measure the number of counts in the selected pixel in every one
of the images. From these numbers, compute a mean count and standard deviation.
Square the standard deviation to get the variance.
-
Use the Signal and Variance from step 6 to add a data point to
your Signal - Variance plot.
-
Change the light intensity and repeat steps 1 through 7.
|
|
|
|
Summary
|
We have derived the mathematical relationship between Signal
and Variance in a CCD image which includes the pixel-to-pixel response
variations among the image pixels. This "flat field" effect must be compensated
or the calculated value of the camera gain will be incorrect. We have shown how
the traditional "simple" method used for gain calculation leads to an erroneous
gain value unless flat field effects are not considered. We have suggested 2
methods that correctly account for the flat field effect, and these should be
implemented in camera testing procedures.
|
|
|
Copyright © 1998. Michael V. Newberry, Mirametrics, Inc. All Rights Reserved.
|
|
|
|
|
|
Previous Mirametrics Article
|
Next Mirametrics Article |
|
|
Opticstar Showroom
|
|
Opening Hours
|
|
updated: 2nd January 2013
|
 |
Day Opening Hours
Monday 9am - 5pm
Tuesday 9am - 5pm
Wednesday 9am - 5pm
Thursday 9am - 5pm
Friday 9am - 5pm
Saturday 10am - 2pm
Sunday CLOSED
We are closed on Sundays and public holidays.
|
|
|
| more |
|
Willmann Bell
|
|
Books for Astronomy
|
|
updated: 15th May 2012
|
 |
|
Acclaimed books from Willmann Bell...
|
|
|
| more |
|
Opticstar News
|
|
Imaging Brochure
|
|
posted: 17th August 2012
|
|
|
Download the Opticstar imaging catalogue here.
|
|
|
| more |
|
Sample Images
|
|
Opticstar PL-130M CoolAir
|
|
posted: 10th February 2009
|
 |
|
An impressive 9.4 megapixel mosaic of the Moon by Herwig Sulzbacher with
the Opticstar PL-130M.
|
|
|
|
Articles
|
|
Pixel Response Effects On CCD Gain Calibration
|
|
|
The gain of a CCD camera is the conversion between...
|
|
|
| more |
|
A&M 152mm f/8 Review
|
|
"Fantastic Performance" "Beauty of a Ferrari"
|
|
updated: 25th March 2009
|
 |
|
Read the latest review about the A&M 152mm f/8 A&M/TMB on
Cloudy Nights.
According to the review this telescope combines the "beauty of a Ferrari" with
"fantastic performance". The review also states that "the manufacturer's claims are valid".
|
|
|
|